Restart both the PC and the AA instrument; check Device Manager. Conclusion
By consulting these resources, users can gain a deeper understanding of the Syngistix for AA software and its applications, and optimize their use of the software to achieve accurate and reliable results. syngistix for aa software manual
| Error Message | Probable Cause | Software Fix (from manual) | | :--- | :--- | :--- | | "Lamp Energy Low" | Lamp misaligned or aged | Run the "Lamp Optimization" routine. Check lamp current settings. | | "Flame Out" | Gas pressure drop or leak | Check gas regulators; click "Reset Flame" in the Safety Panel. | | "Furnace Over-temp" | Cooling water flow failure | Halt run; check chiller water flow rate in the System Status tab. | | "Autosampler Position Failed" | Obstructed rack or motor stall | Use "Jog" controls to manually move the arm; re-home the autosampler. | Restart both the PC and the AA instrument;